○ Ultrafast response time, τ ~ 20 msec
○ Small particle detection down to 1.9 nm
○ Integrated, compact design
○ Patent-pending
○ Aerosol flow path facilitates axisymmetric aerosol sheathing and coupling to a focusing nozzle for optical detection.
○ 1 year warranty
○ Monitors particles ≥1.9 nm
○ Rapid response time to detect changes in particle contamination levels.
○ The response is not dependent on particle refractive index or particle shape.
○ Portable and easy to set up. Robust with minimal maintenance requirements.
○ Simple to operate using the intuitive front-panel-display menus.
|
FCPC 3650 |
---|---|
Particle size detection | 1.9 nm to >3 µm |
1.9 nm to >3 µm | 1 to 100,000 particles/cm3 |
Response time | 50% response time ~ 80 ms, 10-90% response time ~ 35 ms, time constant (τ) ~ 20 ms |
Working condensing fluid | n-butyl alcohol |
Flow control | Critical orifices with internal pumps |
Aerosol sample flow | 300 cm3/min |
Inlet flow | 600 or 1500 cm3/min (user selectable) Aerosol inlet Front panel, or vertical from the bottom (user selectable, fitting access on right side of instrument) |
Dimensions LWH cm (inches) | L: 21.6 cm ×W: 19 cm × H: 21.6 cm) |
Weight | 15 lbs (6.8 kg) |
Power requirements | 50/60 Hz, 100-220 VAC, 75 Watts |
I/O | RJ-45 with Ethernet, 9-pin D-subminiature connector with RS-232 serial communications, pulse output and user selectable analog output |